Section 3 Calibrations

The following section includes all the calibrations in the 2600.

3.1 Power Supply Adjustment

NOTE: THIS ADJUSTMENT MUST BE PERFORMED PRIOR TO MAKING ANY OTHER CALIBRATIONS.

1. Connect a Digital Voltmeter across the plus 15.00 connection and ground (red and black wire).

2. Adjust R10 for exactly +15.00VDC.

3. Connect the DVM across the negative supply connection and ground (black and violet).

4. Adjust R13 for exactly -15.00VDC.

3.2 2600 Tuning