Manual Supplement

Manual Title: Fluke 192B/196B-C/199B-C

Service Manual Supplement Issue: 3

Part Number: 4822 872 05391 Part Number: 4822 872 08633

Print Date: September 2002 Issue Date: 07-April-05

Revision/Date: 0 Page Count 2

This supplement contains information necessary to ensure the accuracy of the above manual.

© 2005 Fluke Corporation. All rights reserved. Printed in the Netherlands

MSU205391.0-3

Fluke 192B/196B-C/199B-C Service Manual

Manual Supplement

The following changes must be made to the service manual:

Chapter 4, section 4.8.3 Continuity Function Test

Replace step 3 with the following:

3. Set the 5500A to 20 Q. Use the 5500A "COMP OFF" mode

Chapter 4, section 4.8.4 Diode Test Function Test

Replace step 3 with the following:

4. Set the 5500A to 1 kQ. Use the 5500A "COMP OFF" mode

Chapter 8

Page

Item

Change

Remark

8-8

C1027

CC 1PF 5% 0805 NP0 50V

4022 301 60051

Added from PCA rev. level 12

8-9

C1144

Cap 68UF 20% 6.3V NBO CASE-C must be

4022 301 63731

Cap 100UF 20% 6.3V NBO CASE-C

4022 301 61211

use this ordering code for all NEW PCA's

8-9

C1227

CC 1PF 5% 0805 NP0 50V

4022 301 60051

Added from PCA rev. level 12

8-10

C1344

Cap 22UF 6.3V 10% X5R 1210 must be

4022 101 00021

Cap 100UF 20% 6.3V NBO CASE-C

4022 301 61211

use this ordering code for all NEW Mainboard PCA's

8-11

C1577

TACAP 10V SMD 20% 100 UF CAP 100UF 10% 10V SMD MNR must be

4022 301 61231 4022 301 61531

SMD CAP 100UF 10% 16V

4022 301 62941

use this ordering code for all OLD and NEW PCA's

8-11

C1579

TACAP 10V SMD 20% 100 UF CAP 100UF 10% 10V SMD MNR must be

4022 301 61231 4022 301 61531

SMD CAP 100UF 10% 16V

4022 301 62941

use this ordering code for all OLD and NEW PCA's

8-14

C3512

C3512 is not placed on the NEW main PCA

8-15

C4008

CAP 100UF 20% 6.3V NBO CASE-D must be

4022 101 00201

CAP 68UF 20% 6.3V NBO CASE-D

4022 101 63731

use this ordering code for all NEW PCA's

8-17

L1301

CHIP INDUCT. 1UH 10% must be

5322 157 63648

FERRITE BEAD 0E

4330 030 35851

use this ordering code for all OLD and NEW PCA's

8-17

L4010

CHIP INDUCT. 47UH 10% must be

4822 157 70794

CHIP INDUCT 330UH

4022 104 00491

Changed from PCA rev. level 11, can be used in all PCA versions

8-18

L4020

CHIP INDUCT. 1UH 10%

5322 157 63648

Added from PCA rev. level 11

8-18

N1576

installed type can also be TC595002ECBTR

Ordering code not changed

8-28

V4004, V4014, V4025

Installed type can also be BYM07-200

Ordering code not changed

8-29

X3601

DISPLAY CONNECTOR 22-P must be

DISPLAY CONNECTOR 22-P

4022 303 10571 4022 303 10501

Listed number is wrong

Manual Supplement

Fluke 192B/196B-C/199B-C Service Manual

Chapter 5.6

The line before ERROR MESSAGES indicates the wrong key ( [ ) ). Please change this line: 'It is allowed to repeat a step that shows the status: READY by pressing (^HjjJ again.'

Chapter 9, Figure 9-9 and Figure 9-14

Add L4020, see the figures below.

L4020 is located in series with V4015 in the +5V2 supply on the Power Circuit. On the PCA one side of V4015 is lifted and L4020 is mounted between the lifted side and the solder spot that became free.

V4015 L4020

L4015 ^vwv.

C4015

C4016

Ceronix 1493

Position of L4020 in Fig. 9-9 location D1

Position of L4020 in Fig. 9-14 location D2

Table

Was this article helpful?

0 0
DIY Battery Repair

DIY Battery Repair

You can now recondition your old batteries at home and bring them back to 100 percent of their working condition. This guide will enable you to revive All NiCd batteries regardless of brand and battery volt. It will give you the required information on how to re-energize and revive your NiCd batteries through the RVD process, charging method and charging guidelines.