411 DC and AC Amps Calibration Front Panel

The dc and ac amps calibration requires a boost amplifier or transconductance amplifier capable of 10 A output at both dc and 1 kHz.

1. If the meter is already in calibration mode, press the (^B or button to select A DC calibration. Otherwise, select calibration mode first (press Cal Enable button for three seconds), then select A DC calibration by pressing the ( button four times.

2. Connect the output of the boost amplifier to the [10 A] and COM inputs of the meter.

3. Apply the dc amp output indicated in Table 4-5 (step 22), and calibrate using the A button.

4. Select A AC calibration on the meter by pressing the ( button once.

5. Apply the two ac amp outputs called for in steps 23 and 24 of Table 4-5, calibrating each step by using the (auto) button.

DIY Battery Repair

DIY Battery Repair

You can now recondition your old batteries at home and bring them back to 100 percent of their working condition. This guide will enable you to revive All NiCd batteries regardless of brand and battery volt. It will give you the required information on how to re-energize and revive your NiCd batteries through the RVD process, charging method and charging guidelines.