43 Performance Tests

The following performance tests are provided to ensure that the meter is in proper operating condition. If the meter fails any of the performance tests, calibration adjustment and/or repair is needed. To perform these tests, you will need a Fluke 5700A Multifunction Calibrator and a 5725A Amplifier or equipment meeting the minimum specifications given in Table 4-1.

Each of the measurements listed in the following steps assumes the meter is being tested after a one-hour warmup in an environment with an ambient temperature of 18 to 28°C and a relative humidity of less than 90% (70% for MW ranges). The limits in Table 4-2 represent the 1-year calibration cycle.

Note

All measurements listed in the performance test tables are made in the medium reading rate unless otherwise noted.

1. Power up the meter and allow it to stabilize for one hour.

2. Connect a cable from the Output VA HI and LO connectors of the 5700A to the Vn^^ and COM connectors on the Fluke 45. Select the function and range on the Fluke 45 and the input level from the 5700A using the values listed in Table 4-2. The display should read between the minimum and maximum values listed in the table.

3. Connect a cable from the Output VA HI and LO connectors of the 5700A to the [100 mA] and COM connectors on the Fluke 45. Select the function and range on the Fluke 45 and the input level from the 5700A using the values contained in Table 4-3. The display should read between the minimum and maximum readings listed in the table.

4. The following tests require a Fluke 5725A Amplifier (or equivalent) to be used with the 5700A. The input level for the performance test will be set on the 5700A but will be output from the 5725A Amplifier.

Connect a cable from the Output VA HI and LO connectors of the 5725A to the [10 A] and COM connectors on the Fluke 45. Select the function and range on the Fluke 45 and the input level from the 5700A using the values contained in Table 4-4. The display should read between the minimum and maximum readings listed in the table.

Table 4-2. Performance Tests for Volts, Diode Test, Ohms, and Frequency Functions

Function

Range (Rate)***

Input Level

Frequency

Display

Min

Max

100 mV (slow)

short

-

-0.006

0.006

90 mV

-

89.971

90.029

1000 mV (slow)

900 mV

-

899.71

900.29

300 mV

short

-

-0.02

0.02

300 mV

-

299.90

300.10

3 V

-3 V

-

-3.0010

-2.9990

30 V

-30 V

-

-30.010

-29.990

300 V

-300 V

-

-300.10

-299.90

1000 V

-1000 V

-

-1000.5

-999.5

30 V

30 V

-

29.990

30.010

300 V

300 V

-

299.90

300.10

1000 V

1000 V

-

999.5

1000.5

(q

-

short

-

-0.0008

0.0008

(tone)

(tone)

-

open

-

OL

OL

300 mV

short

-

<.075

15 mV

1 kHz

14.87

15.13

100 kHz

13.75

16.25

300 mV

1 kHz

299.30

300.70

100 kHz

284.50

315.50

3 V

3.0 V

1 kHz

2.9930

3.0070

30 V

30 V

1 kHz

29.930

30.070

300 V

300 V

1 kHz

299.30

300.70

750 V

750 V

1 kHz

747.5

752.5

y

Using decades of 3:

300 W*

short

0.00

0.04

300 W

299.83

300.19

3 kW*

short

0.000

0.0002

3 kW

2.9983

3.0017

30 kW

30 kW

29.983

30.017

300 kW

300 kW

299.83

300.17

3 MW

3 MW

2.9980

3.0020

30 MW**

30 MW

29.922

30.078

300 MW

300 MW

294.0

306.0

Using decades of 1.9:

300 W*

short

0.00

0.04

190 W

189.88

190.14

Table 4-2. Performance Tests for Volts, Diode Test, Ohms, and Frequency Functions (cont)

Function

Range (Rate)***

Input

Frequency

Display

Level

Min

Max

y

3k W*

short

0.0000

0.0002

1.9 kW

1.8988

1.9012

30 kW

19 kW

18.988

19.012

300 kW

190 kW

189.88

190.12

3 MW

1.9 MW

1.8987

1.9013

30 MW**

19 MW

18.949

19.051

300 MW**

190 MW

186.2

193.8

Using decades of 1.0:

300 W*

short

0.00

0.04

100 W

99.93

100.09

3 kW

short

0.0000

0.0002

1 kW

0.9993

1.0007

30 kW

10 kW

9.993

10.007

300 kW

100 kW

99.93

100.07

3 MW

1 MW

0.9992

1.0008

30 MW**

10 MW

9.972

10.028

300 MW**

100 MW

98.0

102.0

[FREQj

10/100 kHz

0.1 to 10 V

10 kHz

9.994

10.006

* Use either 2-wire compensation on the 5700A or the relative (REL) mode on the 45.

** Optional test points that can be used if standards are available.

*** All measurement rates are in medium speed unless otherwise specified.

Table 4-3. Performance Tests for mA Current Functions

Function

Range (Rate)

Input Level

Frequency

Display

Min

Max

(gl (mA)

30 mA 100 mA

+30 mA + 100 mA

-

29.982 99.93

30.018 100.07

(hi (mA)

30 mA 100 mA

30 mA 100 mA

1 kHz 1 kHz

29.840 99.40

30.160 100.60

Table 4-4. Performance Tests for Current Functions

Function

Range (Rate)

Input Level

Frequency

Display

Min

Max

g

10 A

+ 10 A

-

9.975

10.025

h

10 A

10 A

1 kHz

9.890

10.110

DIY Battery Repair

DIY Battery Repair

You can now recondition your old batteries at home and bring them back to 100 percent of their working condition. This guide will enable you to revive All NiCd batteries regardless of brand and battery volt. It will give you the required information on how to re-energize and revive your NiCd batteries through the RVD process, charging method and charging guidelines.