49 AC Volts Calibration Front Panel

To perform ac volts calibration, proceed as follows:

1. If the meter is already in calibration mode, press the (^B or button to select VAC calibration. Otherwise, select calibration mode first (press Cal Enable button for three seconds), then select VAC calibration by pressing the ( button once.

2. Connect the ac source to the V and COM inputs of the meter.

3. For each of steps 11 through 17 (Table 4-5), apply the source output indicated and calibrate the meter by pressing the A button.

DIY Battery Repair

DIY Battery Repair

You can now recondition your old batteries at home and bring them back to 100 percent of their working condition. This guide will enable you to revive All NiCd batteries regardless of brand and battery volt. It will give you the required information on how to re-energize and revive your NiCd batteries through the RVD process, charging method and charging guidelines.