DC Voltage

Resolution

Accuracy

Range

Slow

Medium

Fast

(6 months)

(1 year)

300 mV

-

10 |V

100 |V

0.02% + 2

0.025%+2

3 V

-

100 |V

1 mV

0.02% + 2

0.025% +2

30 V

-

1 mV

10 mV

0.02% + 2

0.025%+2

300 V

--

10 mV

100 mV

0.02% + 2

0.025% +2

1000 V

-

100 mV

1 V

0.02% + 2

0.025%+2

100 mV

1 |JV

-

-

0.02% + 6

0.025%+6

1000 mV

10 |V

-

-

0.02% + 6

0.025% +6

10 V

100 |V

-

-

0.02% + 6

0.025%+6

100 V

1 mV

-

-

0.02% + 6

0.025% +6

1000 V

10 mV

-

-

0.02% + 6

0.025%+6

Input Impedance

10 MW in parallel with <100 pF

Note

In the dual display mode, when the volts ac and volts dc functions are selected, the 10 MW dc input divider is in parallel with the 1 MW ac divider.

Normal Mode Rejection Ratio

>80 dB at 50 or 60 Hz, slow and medium rates

>54 dB for frequencies between 50-440 Hz, slow and medium rates

>60 dB at 50 Hz, fast rate (Note: Fast rate has no filtering)

Maximum Allowable AC Voltage While Measuring DC Voltage

Peak Normal Mode Signal

Range

Max Allowable Peak AC Voltage

NMRR* >80dB**

NMRR >60 dB**

300 mV

100 mV

20 V

15 V

15 V

3 V

1000 mV

20 V

15 V

15 V

30 V

10 V

1000 V

50 V

300 V

300 V

100 V

1000 V

50 V

300 V

1000 V

1000 V

1000 V

200 V

1000 V

* NMRR is the Normal Mode Rejection Ratio ** Normal Mode Rejection Ratio at 50 or 60 Hz ± 0.1%

Common Mode Rejection Ratio

>90 dB at dc, 50 or 60 Hz, (1 kW unbalanced, medium and slow rates)

Maximum Functional Input

1000 V dc or peak ac on any range

DIY Battery Repair

DIY Battery Repair

You can now recondition your old batteries at home and bring them back to 100 percent of their working condition. This guide will enable you to revive All NiCd batteries regardless of brand and battery volt. It will give you the required information on how to re-energize and revive your NiCd batteries through the RVD process, charging method and charging guidelines.