Performance Conditions

The following electrical characteristics (Table 1-1) are valid for the 2213 when it has been adjusted at an ambient temperature between +20°C and +30°C, has had a warm-up period of at least 20 minutes, and is operating at an ambient temperature between 0°C and +50°C (unless otherwise noted).

Items listed in the "Performance Requirements" column are verifiable qualitative or quantitative limits, while items listed in the "Supplemental Information" column are either explanatory notes, calibration setup descriptions, performance characteristics for which no absolute limits are specified, or characteristics that are impractical to check.

Environmental characteristics are given in Table 1-2. The 2213 meets the requirements of MIL-T-28800B, Class 5 equipment, except where otherwise noted.

Physical characteristics of the instrument are listed in Table 1-3.

Table 1-1 Electrical Characteristics

Characteristics

Performance Requirements

Supplemental Information

VERTICAL DEFLECTION SYSTEM

Deflection Factor Range

2 mV per division to 10 V per division in a 1-2-5 sequence.

1X gain adjusted with VOLTS/DIV switch set to 20 mV per division.

10X gain adjusted with VOLTS/DIV switch set to 2 mV per division.

Accu racy

+20° C to +30° C

±3%.

0°C to +50°C

±4%.a

Range of VOLTS/DIV Variable Control.

Continuously variable between settings. Increases deflection factor by at least 2.5 to 1.

Step Response Rise Time

Measured with a vertically centered 5-division reference signal from a 50-12 source driving a 50-12 coaxial cable that is terminated in 50 12 at the input connector, with the VOLTS/DIV Variable control in its CAL detent.

Rise time is calculated from the formula:

20 mV to 10 V per Division

Dc to at least 60 MHz.

Measured with a vertically centered 6-division reference signal from a 50-12 source driving a 50-12 coaxial cable that is terminated in 50 12, both at the input connector and at the P6120 probe input, with the VOLTS/DIV Variable control in its CAL detent.

2 mV to 10 mV per Division

Dc to at least 50 MHz.

2 mV to 10 V per Division

Dc to at least 50 MHz.a

Chop Mode Repetition Rate

250 kHz ±30%.

Table 1-1 (cont)

Characteristics

Performance Requirements

Supplemental Information

VERTICAL DEFLECTION SYSTEM (cont)

Input Characteristics Resistance

1 MS2 ±2%.a

Capacitance

30 pF ±3 pF.a

Maximum Safe Input Voltage ^ DC Coupled

400 V (dc + peak ac) or 800 V p-p ac to 1 kHz or less.3

AC Coupled

400 V (dc + peak ac) or 800 V p-p ac to 1 kHz or less.3

Common-Mode Rejection Ratio (CMRR)

At least 10 to 1 at 10 MHz.

Checked at 20 mV per division for common-mode signals of 8 divisions or less, with VOLTS/DIV Variable control adjusted for best CMRR at 50 kHz.

TRIGGER SYSTEM

Trigger Sensitivity AUTO and NORM

0.4 division internal or 50 mV external to 2 MHz, increasing to 1.5 divisions internal or 250 mV external at 60 MHz.

External trigger signal from a 50-£2 source driving a 50-£2 coaxial cable that is terminated in 50 at the input connector.

Will trigger on tv line sync components in NORM only: > 0.4 division internal or 50 mV p-p external.

AUTO Lowest Usable Frequency

20 Hz.3

TV FIELD

2.0 divisions of composite video or composite sync.3

External Input

Maximum Input Voltage ^

400 V (dc + peak ac) or 800 V p-p ac at 1 kHz or less.3

Input Resistance

1 Mfi ±2%.3

Input Capacitance

30 pF ±3 pF.a

AC Coupled

10 Hz or less at lower —3 dB point.3

LEVEL Control Range

(with NORM TRIGGER MODE)

INT

On screen limits.3

EXT and DC

At least ±2 V (4 V p-p).3

Table 1-1 (cont)

Characteristics

Performance Requirements

Supplemental Information

TRIGGER SYSTEM (cont)

LEVEL Control Range (with NORM TRIGGER MODE) (cont)

EXT and DC -M0

At least ±20 V (40 V p-p).a

VAR HOLDOFF Control Range

Increases sweep holdoff time by at least a factor of four.3

HORIZONTAL DEFLECTION SYSTEM

HORIZONTAL DEFLECTION SYSTEM

Sweep Rate

Calibrated Range A Sweep

0.5 s per division to 0.05 fis per division in a 1-2-5 sequence. X10 Magnifier extends maximum sweep speed to 5 ns per division.

Accu racy

Unmagnified

Magnified

POSITION Control Range

Start of sweep to 100th division will position past the center vertical graticule line with X10 Magnifier.

Variable Control Range

Continuously variable between calibrated settings. Extends the sweep speeds by at least a factor of 2.5.

Delay Time

Range Selector

MULTIPLIER Control

Jitter

Minimum delay is less than selected values of 0.5 jus, 10 fis, and 0.2 ms.

Increases delay time by at least a factor of 20.

One part, or less, in 5,000 (0.02%) of the maximum available delay time.

Sweep accuracy applies over the center 8 divisions. Exclude the first 25 ns of the sweep for both magnified and unmagnified sweep speeds and exclude anything beyond the 100th magnified division.

X-Y OPERATION (X1 MAGNIFICATION)

Deflection Factors

Range

Same as Vertical Deflection System, with both VOLTS/DIV Variable controls in CAL detent.

Specification—2213 Service

Table 1-1 (cont)

Characteristics

Performance Requirements

Supplemental Information

X-Y OPERATION (X1 MAGNIFICATION) (cont)

Deflection Factors (cont) Accuracy

+20°C to +30°C

X-Axis

Y-Axis

Measured with a dc-coupled, 5-division reference signal.

±5%

±3%

0°C to +50°C

+ 6%a

±4%a

Bandwidth X-Axis

Dc to at least 2 MHz.

Measured with a 5-division reference signal.

Y-Axis

Same as Vertical Deflection System.

Phase Difference Between X- and Y-Axis Amplifiers

±3° from dc to 50 kHz.®

With dc-coupled inputs.

PROBE ADJUST

Signal at PROBE ADJUST Jack Voltage

0.5 V ±20%.

Repetition Rate"

1 kHz ±20%.a

Z-AXIS INPUT

Sensitivity

5 V causes noticeable modulation. Positive-going input signal decreases intensity

Usable Frequency Range

Dc to 5 MHz.3

Maximum Safe Input Voltage

30 V (dc + peak ac) or 30 V p-p ac at 1 kHz or less.3

Input Impedance

10 kS2 ±10%.a

POWER SOURCE

Line Voltage Range

90V to 250V.a

Line Frequency Range

48 Hz to 62 Hz.a

Instruments with Current Limit Board

48 Hz to 440 Hz.a

Instruments with Preregulator Board

Maximum Power Consumption

50 W.a

Instruments with Current Limit Board

40 W.a

Instruments with Preregulator Board

Line Fuse

2A, 250V, Fast

Instruments with Current Limit Board

1 A, 250V, Slow

Instruments with Preregulator Board

Table 1-1 (cont)

Characteristics

Performance Requirements

Supplemental Information

CATHODE-RAY TUBE

Display Area

80 by 100 mm.a

Standard Phosphor

P31.a

Nominal Accelerating Voltage

10,000 V.a

Performance Requirement not checked in Service Manual.

Performance Requirement not checked in Service Manual.

Table 1-2 Environmental Characteristics

Characteristics

Description

The instrument meets all of the following MIL-T-28800B requirements for Class 5 equipment.

Temperature Operating

0°C to +50°C (+32° F to +122°F).

Nonope rating

—55°C to +75°C (—67°F to +167°F).

Altitude Operating

To 4,500 m (15,000 ft). Maximum operating temperature decreased 1°C per 300 m (1,000 ft) above 1,500 m (5,000 ft).

Nonoperating

To 15,000 m (50,000 ft).

Humidity (Operating and Nonoperating)

5 cycles (120 hours) referenced to MIL-T-28800B, Class 5 instruments.

Vibration (Operating)

15 minutes along each of 3 major axes at a total displacement of 0.015 inch p-p (2.4 g at 55 Hz), with frequency varied from 10 Hz to 55 Hz to 10 Hz in 1-minute sweeps. Hold for 10 minutes at 55 Hz. All major resonances must be above 55 Hz.

Shock (Operating and Nonoperating)

30 g, half-sine, 11-ms duration; 3 shocks per axis each direction, for a total of 18 shocks.

Table 1-3 Physical Characteristics

Characteristics

With Front-Panel Cover, Accessories, and Pouch

7.6 kg (16.81b).

Without Front-Panel Cover, Accessories, and Pouch

6.1 kg (13.5 lb).

Domestic Shipping

8.2 kg (18.0 lb).

Height With Feet and Handle

With Handle

361 mm (14.2 in).

Without Handle

328 mm (12.9 in).

Depth

With Front-Panel Cover

445 mm (17.5 in).

Without Front-Panel Cover

439 mm (17.3 in).

With Handle Extended 511 mm (20.1 in).

With Handle Extended 511 mm (20.1 in).

0 0

Post a comment