Preparation

Test equipment items 1 through 9 in Table 4-1 are required to accomplish a complete Performance Check. At the beginning of each subsection, in both the "Performance Check Procedure" and the "Adjustment Procedure" sections, there is an equipment-required list showing only the test equipment necessary for performing the steps in that subsection. In this list, the item number that follows each piece of equipment corresponds to the item number listed in Table 4-1.

This procedure is structured in subsections, which can be performed independently, to permit checking individual portions of the instrument. At the beginning of each subsection is a list of all the front-panel control settings required to prepare the instrument for performing Step 1 in that subsection. Each succeeding step within a particular subsection should then be performed, both in the sequence presented and in its entirety, to ensure that control-setting changes will be correct for ensuing steps.

Table 4-1 Test Equipment Required

Item No. and Description

Minimum Specification

Purpose

Examples of Suitable Test Equipment

1. Calibration Generator

Standard-amplitude signal levels: 10 mV to 50 V. Accuracy: ±0.3%.

High-amplitude signal levels: 1 V to 60 V. Repetition rate: 1 kHz.

Fast-rise signal level: 1 V. Repetition rate: 1 MHz. Rise time: 1 ns or less. Flatness: ±0.5%.

Vertical and horizontal checks and adjustments.

TEKTRONIX PG 506 Calibration Generator.3

2. Leveled Sine-Wave Generator

Frequency: 250 kHz to above 70 MHz. Output amplitude: variable from 10 mV to 5 V p-p. Output impedance: 50 £2. Reference frequency: 50 kHz. Amplitude accuracy: constant within 3% of reference frequency as output frequency changes.

Vertical, horizontal, and triggering checks and adjustments. Display adjustment and Z-axis check.

TEKTRONIX SG 503 Leveled Sine-Wave Generator.3

3. Time-Mark Generator

Marker outputs: 10 ns to 0.5 s. Marker accuracy: ±0.1%. Trigger output: 1 ms to 0.1 jus, time-coincident with markers.

Horizontal checks and adjustments. Display adjustment.

TEKTRONIX TG 501 TimeMark Generator.3

4. Cable (2 required)

Impedance: 50 £2. Length: 42 in. Connectors: bnc.

Signal interconnection.

Tektronix Part Number 012-0057-01.

5. Termination (2 required)

Impedance: 50 £2. Connectors: bnc.

Signal termination.

Tektronix Part Number 011-0049-01.

6. Dual-Input Coupler

Connectors: bnc-female-to-dual-bnc male.

Vertical checks and adjustments.

Tektronix Part Number 067-0525-01.

7. 10X Attenuator

Ratio: 10X. Impedance: 50£2. Connectors: bnc.

Vertical compensation and triggering checks.

Tektronix Part Number 011-0059-02.

8. T-Connector

Connectors: bnc.

Signal interconnection.

Tektronix Part Number 103-0030-00.

9. Adapter

Connectors: bnc-male-to-miniature probe tip.

Signal interconnection.

Tektronix Part Number 013-0084-02.

10. Variable Autotransformer

Capable of supplying 1.5 A at 115 V.

Instrument input voltage adjustment.

General Radio W8MT3VM Variac Autotransformer.

aRequires a TM 500-series power-module mainframe.

aRequires a TM 500-series power-module mainframe.

Table 4-1 (cont)

Item No. and Description

Minimum Specification

Purpose

Examples of Suitable Test Equipment

11. Digital Voltmeter

Range: 0 to 140 V. Dc voltage accuracy: ±0.15%. 4 1/2-digit display.

Power supply checks and adjustment.

Vertical adjustment.

TEKTRONIX DM 501A Digital Multimeter.8

12. Test Oscilloscope with included 10X probe (Standard Accessory) and 1X probe (1X probe is option^accessory).

Bandwidth: dc to 10 MHz. Minimum deflection factor: 5 mV/div. Accuracy: ±3%.

Power supply ripple check and general troubleshooting.

a. TEKTRONIX 2213 Oscilloscope.

b. TEKTRONIX P6101 Probe (IX). Part Number 010-6101-03.

13. DC Voltmeter

Range: 0 to 2500 V, calibrated to 1% accuracy at -2000 V.

High-voltage power supply check.

Triplett Model 630-NA.

14. Screwdriver

Length: 3-in shaft. Bit size: 3/32 in.

Adjust variable resistors.

Xcelite R-3323.

15. Low-Capacitance Alignment Tool

Length: 1-in shaft. Bit size: 3/32 in.

Adjust variable capacitors.

J.F.D. Electronics Corp. Adjustment Tool Number 5284.

aRequires a TM 500-series power-module mainframe.

aRequires a TM 500-series power-module mainframe.